The MultiDetek3 is the latest development of our popular range of compact industrial gas chromatographs. It is more than a simple gas chromatograph: it’s a modular process GC instrument, that combines the functionality of two GCs in one with the ability to also monitor trace moisture and oxygen online.

The instrument has been completely redesigned and features an extra-large 15.6” touch screen for easy visibility and operation. It has a dual sample inlet and online modules which allow simultaneous stream analysis of trace moisture and oxygen together with trace impurities. It is temperature controlled for improved stability and performance. The software interface is our latest LDChroma+, which offers more sophisticated data management and a planning platform.

THE MULTIDETEK 3 SPECIFICATION

GAS CHROMATOGRAPHY DETECTORS
  • PED / TCD / FID
ONLINE DETECTORS
  • Quartz Crystal Microbalance / Ceramic Metal-Oxide / Electrochemical / Zirconia
STANDARD FEATURES
  • Temperature controlled process GC
  • 15,6’’ wide touchscreen with 1366 x 768 HD resolution (IP65)
  • Windows 10 IoT
  • Ethernet ports available for remote control and Modbus
  • Self diagnostic system with maintenance planning
  • Isothermal and/or programmable ramping ovens
  • Electronic flow control regulators for carrier & sample gases
  • 4-20 mA isolated outputs
  • Alarm Historic
  • Digital system status output for remote monitoring (dry relay contact)
  • 2 alarms contact
  • High resolution Chromatogram output
Options
  • Serial communication (RS232/485) / Profibus / Modbus / Ethernet
  • Compact purifier attached to the chassis for generating high purity carrier gas
  • Integrated stream selector system
  • Digital inputs for remote starting
  • Analog inputs for connecting external instruments
  • Remote control for stream selector (LDGSS)
  • Split/split less injector (can be heated)
  • Dual sample inlets with simultaneous analysis
  • Built in online sensor module for trace moisture: Quartz Crystal (ppb/ppm) or Ceramic (ppm)
  • Built in online sensor module for trace oxygen: Electrochemical (ppm) or Zirconia (ppm)
GAS CONNECTIONS
  • 1/8’’ or 1/4’’ VCR or compression fittings
CARRIER PRESSURE REQUIREMENTS
  • 100 PSIG (other carrier pressure available on request)
AMBIENT OPERATING TEMPERATURE RANGE
  • 10 °C - 45 °C
AMBIENT TEMPERATURE CONTROL RANGE FEATURE
  • 20°C - 40°C (out of this range, surrounding ambient temperature must be stabilized)
SUPPLY
  • 115 VAC, 50 – 60 Hz or 220 VAC, 50 – 60 Hz
POWER CONSUMPTION
  • Maximum 500W
REPEATABILITY
  • Three times the percentage of deviation (3*CV %) of each component has to be smaller than 5% on ten consecutive cycles
LDL
  • 3 times noise level
LOQ
  • 3 times LDL value
LONG TERM STABILITY & DRIFT
  • Three times the percentage of deviation (3*CV %) of each component has to be smaller than 10% for 8 hours
LINEARITY
  • For 5 points within the measuring range, the linear curve must have its R2 at a value between 0.998 and 1.00